Niger J Paed 2015; 42 (1): 34 – 38
ORIGINAL
Oyenusi EE
Reliability of bedside blood
Oduwole AO
Oladipo OO
glucose estimating methods in
Njokanma OF
detecting hypoglycaemia in the
Esezobor CI
children’s emergency room
DOI:http://dx.doi.org/10.4314/njp.v42i1,8
Accepted: 17th October 2014
Abstract: Background: Hypogly-
0.05) between Accuchek Active®
caemia occurs in many disease
results and laboratory values than
Oyenusi EE
(
)
states common in the tropics. Fa-
was obtained with Betachek Vis-
P.O.Box 4337, Ikeja,
cilities and skilled manpower re-
ual®(r = 0.48, p=0.000).In detec-
Lagos, Nigeria.
quired for laboratory blood glu-
tion of hypoglycaemia, both bed-
Email: ebikike@yahoo.com
cose measurement are not always
side glucose monitors were found
Oduwole AO, Esezobor CI
available in health facilities in
to have a high specificity and high
Department of Paediatrics,
developing countries.
predictive values of a negative test
Lagos University Teaching Hospital,
Objective: The study was carried
(99.8% and 98.5% for Accuchek
Idi-Araba, Lagos, Nigeria.
out to determine the validity of
Active®and 89.4% and 97.8% for
bedside methods of blood glucose
Betachek Visual®respectively)
Oladipo OO
measurement in detecting hypo-
with moderate sensitivity (75.0%
Staten Island University Hospital,
glycaemia.
and 66.7% respectively). However,
Staten Island. NY. USA
Methods: Blood glucose was de-
the Accuchek Active® monitor has
termined by two bedside methods
a much higher predictive value of
Njokanma OF
( Accuchek Active® and Betachek
a positive test (94.1%) compared
Department of Paediatrics,
Visual®) in 430 patients aged
to the Betachek Visual® (27.1%).
Lagos State University Teaching
Hospital, Ikeja, Lagos, Nigeria.
between one month and 10years
Conclusion: The bedside glucose
and simultaneously sent for labo-
monitors are valid bedside tools
ratory spectrophotometric analysis
for detecting or ruling out hypo-
at a wavelength of 500nm using
glycemia.
the hexose kinase method. Hypo-
glycaemia was defined as plasma
Keywords: hypoglycaemia, chil-
glucose < 2.5mmol/L.
dren, bedside investigation, glu-
Results: The prevalence of hypo-
cose estimation, blood glucose.
glycaemia was 5.6%.There was a
higher correlation (r =0.84, p<
Introduction
ever, the absence of clinical symptoms does not always
indicate that the glucose concentration is normal because
hypoglycaemia may be frequently asymptomatic . Thus
8
Hypoglycaemia occurs in many disease states such as
severe malaria, severe malnutrition, diarrhoea amongst
it is important to investigate for hypoglycaemia in
others common in the tropics
1-4
and may also complicate
acutely ill children, more so as it complicates paediatric
treatment with some drugs like quinine used in the treat-
emergencies in Africa
9-11
.
ment of severe malaria . A previous study had docu-
5
3
mented a prevalence rate of 17% among children with
The gold standard technique of determining blood glu-
severe malaria seen at the children’s emergency centre
cose level involves laboratory spectrophotometric analy-
of the Lagos University Teaching Hospital (LUTH) .
3
sis using methods involving enzymes such as glucose
Significant morbidity and mortality can be associated
oxidase, hexokinase and glucose dehydrogenase. These
with hypoglycaemia . Survivors of prolonged hypogly-
6,7
methods require standard laboratory equipment, skilled
caemia are prone to neurological complications such as
manpower, consistent electricity and water supply.
mental retardation, recurrent seizure activity, transient
These resources required for the laboratory determina-
cognitive impairment and neurological deficits .
6,7
tion of blood glucose level are not always available in
our health facilities in the developing countries, yet early
Clinical manifestations of hypoglycaemia include sweat-
diagnosis and prompt treatment of hypoglycaemia in
ing, jitteriness, anxiety and nervousness. Others are
critically-ill children minimizes hypoglycaemia-related
organ damage .
6,7
headaches, visual disturbances, lethargy, restlessness,
irritability, convulsions and mental confusion . How-
8
35
In addition, the cost of laboratory determination of
blood was wiped from the test pads with a folded tissue
blood glucose level could be high when serial measures
paper and repeated using a clean region of the tissue
are required. Because of relatively long turn-around time
paper. It was ensured that strip was blood-free, as excess
blood left on the pads usually affects the result . After
16
results from laboratory determination of glucose may
not be available in a timely fashion to affect manage-
an additional 30 seconds, the test result was read by
ment of a child. To overcome some of these challenges,
comparing the reacted test pad to the B etachek Visual®
various types of reagent test strips have been developed
colour chart label. If the colour fell between two colour
blocks, the average value was taken as the test result .
16
for bedside usage. An added advantage of these rapid
bedside diagnostic tests is that less expertise/training is
required prior to their usage. These strips use a method
For the Accuchek Active® the test strip was held so that
in which a dye is coloured by the glucose oxidase-
the application area and arrows were facing upwards. In
peroxidase chromogenic reaction
12,13
. The resulting col-
this position the test strip was pushed into the glucome-
ter until it clicked into place . The insertion of the test
17
our change is either visually interpreted by comparison
with an accompanying colour chart or read with a reflec-
strip automatically turned the monitor on and put it in
test mode . When the drop symbol flashed in the dis-
17
tance meter (glucometer)
14-15
.
Therefore, it is imperative to validate these bedside
play, a drop of blood was applied to the middle of the
methods for reliability of results obtained from them.
orange coloured test pad. The test was complete after
Hence, this study was conducted to determine the sensi-
about five seconds and the result appeared in the display
tivity, specificity and predictive values of two bedside
and was automatically saved with the date and time. The
methods of glucose determination using a laboratory-
Accuchek Active®monitor measures blood glucose
based enzymatic method (the hexokinase method) as the
within a defined range (10-600 mg/dl or 0.6-33.3 mmol/
gold standard.
L). Values outside this range are displayed as “lo”
meaning the result is less than 0.6 mmol/L or “Hi”
meaning the result is greater than 33.3 mmol/L .
17
The blood in the fluoride oxalate-containing bottle was
Subjects and methods
sent to the Chemical Pathology Department of the hospi-
tal for plasma glucose determination. The blood sample
The study was conducted at the Olikoye Ransome-Kuti
was centrifuged for five minutes at about 1600Gy to
obtain plasma not more than one hour after collection .
18
Children’s Emergency Centre of the (LUTH), Lagos
The separated plasma was frozen at −80 C and stored in
0
Nigeria. All consecutively admitted children, aged be-
tween one month and 10 years over a period of nine
the refrigerator until analysis. Samples were analysed in
months (September 2009 to May 2010) were eligible for
batches of 25 samples using the Hitachi Automatic Ana-
inclusion in the study. Approval was obtained from the
lyser (Cobas- ROCHE). The analyser uses spectropho-
Human Research and Ethics Committee of LUTH before
tometric analysis at a wavelength of 500nm and hexose
kinase method . Hypoglycaemia was defined as whole
12
commencement of the study. Written informed consent
was also obtained from all parents or care –givers of
blood glucose less than 2.2 mmol/L for the bedside
participating children.
methods, ( Betachek Visual® Accuchek Active® ) and as
plasma glucose less than 2.5 mmol/L from the sample
analysed in the laboratory . This was based on the fact
12
Data Collection
that whole blood glucose concentration is approximately
For each subject recruited, at the point of admission into
8 to 15% less than plasma glucose because the water
the emergency room, before commencement of intrave-
content of plasma is approximately 12% higher than that
of whole blood .
12
nous fluids, a drop of capillary blood obtained by a fin-
ger prick was applied to each of the two reagent strips
according to the manufacturer’s instructions
16,17
and
Data management and analysis
about 1.5mls of blood was withdrawn from a convenient
peripheral vein and put in a fluoride oxalate-containing
The data was analysed using the Statistical Package for
bottle for laboratory spectrophotometric analysis using
the Social Sciences (SPSS) version 20. Measures of
the hexose kinase method.
statistical location (mean, standard deviation, median
and range) were determined. Probability (p value) less
A visually interpreted test strip by Betachek Visual and
16
than 0.05 was taken as statistically significant. The sen-
a reflectance meter, Accuchek Active were used for the
17
sitivity, specificity, positive and negative predictive val-
bedside determination of blood glucose because these
ues of blood glucose obtained from the two reagents
are readily available and commonly used in our facili-
strips were calculated using the blood glucose from the
laboratory as the gold standard . For the purposes of
19
ties. The test pad of the Betachek Visual
®
contains
0.09% glucose oxidase, 0.19% peroxidase, 1.2% TMB
this study, the values of sensitivity, specificity, positive
(3, 3', 5, 5'-Tetramethylbenzidine dihydrochloride) and
and negative predictive values of blood glucose were
0.35% DCP (2,4-dichlorophenol 6-monooxygenase) .
16
classified as low if between 0 to <50%, moderate if be-
A test strip was removed and the bottle was immediately
tween 50-80% and high if from >80 to 100%.
recapped. A drop of blood was pressed onto the centre
of two test pads, moving it around to ensure complete
coverage. Exactly 30 seconds after application, the
36
Results
diagnostic accuracy were calculated for results obtained
using the bedside blood glucose monitors (Table 2 for
Four hundred and sixty one children aged 1.5 to 120
Betachek Visual®strips and Table 3 for Accuchek Ac-
months were admitted over a six-month period. Of the
tive®meter). With respect to Betachek Visual®strips,
461 patients, 31 laboratory blood glucose samples
the sensitivity was 66.7% and the specificity was 89.4%.
obtained were not valid due to improper handling. Four
The positive predictive value and the negative predictive
hundred and thirty patients had complete data sets and
value were 27.1% and 97.8% respectively.
were analysed.
Table 2: Sensitivity and specificity of the Betachek Visual®
Blood glucose (BG) patterns of the patients
strip, Laboratory based method
Test
Positive
Negative
Total
The descriptive statistics of blood glucose results by the
Betachek Positive
True Positives = 16 False Positives = 43
59
various methods is shown in Table 1. Using the labora-
Visual ®
tory based method, the mean BG of all the study sub-
Negative False negatives= 8 True Negatives= 363
371
jects was 5.19 ± 2.05 mmol/L (range of 0.1-18.0 mmol/
Total
24
406
430
Sensitivity = TP / (TP + FN) = 16 / (16+8) ×100 = 66.7%
L, median was 4.9 mmol/L).
Specificity = TN / (TN + FP) = 363 / (363+43) ×100 = 89.4%
Positive Predictive Value = TP /(TP+FP)=16/(16+43)×100= 27.1%
Table 1: Descriptive statistics of the blood glucose results by
Negative Predictive Value = TN/(FN+TN)=363/(8+363)×100 = 97.8%
the various methods
Test
Range
Mean
SD
Median
Total
The sensitivity, specificity and the positive and negative
Method
(mmol/L) (mmol/L) (mmol/L) (mmol/L)
predictive values of the Accuchek Active® meterstrips
Laboratory
0.1− 17.2 5.19
2.05
4.9
430
are shown in Table 3. Measures of diagnostic accuracy
Accuchek Active® 0.5 −17.1 5.69
2.04
5.6
430
were calculated for results obtained using Accuchek Ac-
Betachek Visual® 0.0 −18.0 5.54
2.83
6.0
430
tive®meterstrips. The sensitivity was 75.0% while the
specificity was 99.8%.The positive predictive value was
The scatter diagram (Fig 1) demonstrates the relation-
94.7% while the negative predictive value was
ship between laboratory blood glucose and blood glu-
98.7%.Four out of the eight false negatives had blood
cose determined by Accuchek Active® and by Betachek
glucose values ranging from 2.7 to 2.9 mmol/L.
Visual®.Visual inspection further indicates a better
positive relationship between BGs determined by labo-
Table 3: Sensitivity and specificity of the Accuchek Active®
ratory and Accuchek Active® than between laboratory
glucometer, Laboratory based method
and Betachek Visual® .
Test
Positive
Negative
Total
The blood glucose values measured by the Accuchek
Accuchek Positive True Positives =18
False Positives =1
19
Active®meter correlated better with the values from the
Active®
laboratory than those measured by the Betachek Visual®
Negative
False negatives =6 True Negatives = 405 411
strip ( with R, 0.84 versus 0.48) and the difference was
Total
24
406
430
statistically significant (p < 0.05).
Sensitivity= 18/ (18+6) ×100 = 75.0%
Specificity = 405/ (405+1) ×100 = 99.8%
Positive Predictive Value = 18/ (18+1) ×100 = 94.7%
Fig 1: Scatter diagram showing the relationship between Labo-
Negative Predictive Value = 405/ (6+405) ×100 = 98.5%
ratory blood glucose and blood glucose determined by
Accuchek Active®and by Betachek Visual®
Discussion
The blood glucose values measured by the Accuchek
Active®meter correlated better with the values from the
laboratory than those measured by the Betachek Visual®
strip. The chances are very high that an increase/
decrease in laboratory result will be reflected by an in-
crease/decrease in Accuchek Active® meter result.
Correlation of Betachek Visual with laboratory blood glucose, r =
®
0.48, p<0.05
The specificity and negative predictive value of the Be-
Correlation of Accuchek Active with laboratory blood glucose, r=
®
tachek Visual®strip were high, in the range of 90% and
0.84, p<0.05
above. Sensitivity on the other hand was moderate but
Hypoglycaemic patients
positive predictive value was low. This implies that the
use of this technique as a screening tool confers moder-
Twenty-four out of the 430 study subjects had hypogly-
ate success. However, the high specificity and a high
caemia by the laboratory method giving a prevalence
predictive value of a negative test means that a
nega-
rate of hypoglycaemia of 5.6%.
tive result can be relied upon as being truly negative.
Sensitivity and specificity of bedside glucose monitor
The high specificity and negative predictive value of the
for the detection of hypoglycaemia: Using the laboratory
Betachek Visual®strip is comparable to a specificity of
values of blood glucose as gold standard, measures of
94% and negative predictive value of 99% of a different
brand
of
visually-interpreted
reagent
strip
37
( ChemstripbG ) in a previous Cincinnati
14
study. How-
Active®was higher than the 63.1% of the Prestige IQ
ever, the sensitivity and the positive predictive value of
meter. Also, in comparison with findings in a Kenyan
study , the specificity of the Accuchek Active® meter in
20
the Betachek Visual® strip used in the current study was
lower than the reported sensitivity and positive predic-
the current study was higher than the specificity of 48%
tive value of the ChemstripbG strip which was 97% and
but the sensitivity was lower than 84% of the Refloflux-S
72% respectively . A possible explanation for the dis-
14
reflectance meter respectively.
parity may be the definition of hypoglycaemia as labora-
tory blood glucose value of <45 mg/dl (2.5mol/L) in the
The different bedside equipment for measuring blood
current study compared to the higher laboratory blood
glucose are associated with different degrees of success
glucose value of <60mg/dl (3.3mol/L) defined as hypo-
with identifying true positives and true negatives as well
glycaemia in the Cincinnati study . Another possible
14
as avoiding false positives and false negatives. How-
reason for the low sensitivity of the Betachek Visual®
ever, the advantages they offer in terms of low cost,
may be because it is semi-quantitative and measures
immediate availability of results and the option of
blood glucose to the nearest 1mmol/L(18mg/dl) com-
immediate intervention make them very attractive. There
pared to the standard laboratory method which is more
will, however, remain a need for laboratory determina-
quantitative measuring to the nearest 0.1mmol/L(1.8mg/
tion of values, especially in cases where the results are
dl). Furthermore, the visual reading with a colour chart
borderline.
used in the visually interpreted strips may introduce in-
ter-observer bias as people may interprete colour inten-
sity differently while visual acuity and colour blindness
may be sources of bias. However, from the results in the
current study, the Betachek Visual® strip may be more
Conclusion
useful as a confirmatory rather than a screening test for
hypoglycaemia.
In conclusion, the glucose meter, Accuchek Active® and
Betachek visual®are reliable in detecting or ruling out
The specificity and negative predictive value of the
hypoglycaemia by the bedside which will enable prompt
Accuchek Active®were nearly 100%. The positive
intervention while awaiting confirmatory results from
predictive value was also very high but the sensitivity
the laboratory where these are available.
was only moderate. Sensitivity was hampered by the
occurrence of false negative results in six cases but it is
Author's Contributions
noteworthy that four of the six false negative blood glu-
The research work was conceived by all the authors.
cose values were in the lower range of normal. The im-
OEE carried out the field work. All the authors contrib-
plication is that blood glucose values in the lower band
uted to the writing of the manuscript.
of normal may require confirmation using laboratory
Conflict of interest: None
techniques. The high specificity and negative predictive
Funding: None
value imply that the Accuchek Active® meter is very
useful for ruling out hypoglycaemia and a positive value
can be relied upon as being truly hypoglycaemic.
Acknowledgements
The high specificity and negative predictive value of the
Accuchek Active®meter is similar to a specificity of
The authors hereby acknowledge the children and
96.2% and a negative predictive value of 99.7%
parents who participated in this study. We also want to
obtained with a different brand of a blood glucose kit
appreciate all the registrars in the children’s emergency
also interpreted by a reflectance meter ( Prestige IQ ) as
room of the Lagos University Teaching Hospital and the
documented in a previous Nigerian study . However,
15
other research assistants who contributed to the work.
the sensitivity of the Accuchek Active® meter was lower
We also acknowledge the laboratory scientist, Mr Tony
than the sensitivity of 96% of the Prestige IQ meter
Ani for his assistance and guidance in the laboratory
while the positive predictive value of the Accuchek
analysis.
References
1.
Elusiyan JBE, Adejuyigbe EA,
3.
Nwosu SU, Lesi FEA, Mafe AG,
5.
Phillips RE, Looareesuwan S,
Adeodu OO. Hypoglycaemia in a
Egri-Okwaji MTC. Hypoglycae-
White NJ, Chanthavanich P, Karb-
Nigerian paediatric emergency
mia in children with cerebral ma-
wang J, Supanaranond Wet al.
ward. J Trop Paediatr 2006; 25:96
laria seen at the Lagos University
Hypoglycaemia and antimalarial
-102.
Teaching Hospital: predisposing
drugs: quinidine and release of
2.
Osier FHA, Berkley JA, Ross A,
factors. Nig Med J 2004; 45:9-13 .
insulin. BMJ 1986; 292:1319-21.
Sanderson F, Mohammed S, New-
4.
Onyiriuka AN, Awaebe PO,
6.
Idro R, Carter JA, Fegan G,
ton CRJC. Abnormal glucose
Kouyaté M. Hypoglycaemia at
Neville BGR, Newton CRJC. Risk
concentrations on admission to a
point of hospital admission of
factors for persisting neurological
rural Kenyan district hospital:
under-five children with acute
and cognitive impairments follow-
prevalence and outcome. Arch Dis
diarrhoea: prevalence and risk
ing cerebral malaria. Arch Dis
Child 2003; 88:621-5.
factors. Niger J Paed 2013; 40:
Child 2006; 91:142-8.
384-8.
38
16. Betachek Visual test strips
®
7.
Menni F, de Lonlay P, Sevin C,
11. Duru C, Peterside O, Akinbami
Touati G, Peigne C, Barbier V et
F.Pattern and outcome of admis-
[product information] Sidney Aus-
al.Neurological outcomes of 90
sions as seen in the paediatric
tralia: National Diagnostic Prod-
neonates and infants with persis-
emergency ward of the Niger Delta
ucts (NBP) [www.betachek.com];
tent hyperinsulinaemichypogly-
University Teaching Hospital
2006.
17. Accuchek Active blood glucose
®
caemia. Pediatrics 2001; 107:476-
Bayelsa State, Nigeria. Niger J
9.
Paed 2013; 40:232-7.
monitor [user’s manual] Mann-
8.
Thornton TS, Finegold DN,
12. Sacks DB. Carbohydrates. In: Bur-
heim (Germany): Roche Diagnos-
Stanley CA, Sperling MA. Hypo-
tis CA, Ashwood ER (editors).
tics; 2003.
glycaemia in the infant and child.
Tietz Fundamentals of Clinical
18. Cheesebrough M. District Labora-
Chemistry. 5 edition. London:
th
In: Sperling MA (editor) Paediat-
tory Practice in Tropical Countries,
ric Endocrinology. 2 edn. Phila-
nd
WB Saunders; 2001.pp.776-85.
Part 2. Cambridge: Cambridge
delphia: Saunders; 2002.pp.135-
13. Cheesebrough M. District Labora-
University Press; 2000.pp.299.
59.
tory Practice in Tropical Countries,
19. Oyediran ABOO. Screening and
9.
Diakparomre MA, Obi JO. The
Part 2. Cambridge: Cambridge
validity in clinical research. In:
pattern of paediatric emergencies
University Press; 2000.pp.341-3.
Bankole MA (editor). Handbook of
in the University of Benin Teach-
14. Scott PA, Wolf LR, Spadafora MP.
Research Methods for Doctors and
ing Hospital. Niger J Paed 1980;
Accuracy of reagent strips in de-
Dentists. Lagos: National Post-
7:43-5.
tecting hypoglycemia in the emer-
graduate Medical College of Nige-
10. Oviawe O. Comparative study of
gency department. Ann Emerg
ria; 1989.pp.384-9.
the pattern and severity of child-
Med 1998; 32:305-9
20. English M, Wale S, Binns G,
hood diseases seen at the chil-
15. Elusiyan JBE, Adeodu OO, Ade-
Mwangi I, Sauerwein H, Marsh K.
dren’s emergency room of the
juyigbe EA. Evaluating the valid-
Hypoglycaemia on and after ad-
Lagos University Teaching Hospi-
ity of a bedside method of detect-
mission in Kenyan children with
tal in 1967 and in 1982. Niger J
ing hypoglycaemia in children.
severe malaria. Q J Med 1988;
Paed 1987; 14:41-4.
Pediatr Emerg Care 2006; 22:488-
91:191-7.
90.